ENGLISH
La vitrine de diffusion des publications et contributions des chercheurs de l'ÉTS
RECHERCHER

Axial tension/compression and torsional loading of diamond and gyroid lattice structures for biomedical implants: Simulation and experiment

Téléchargements

Téléchargements par mois depuis la dernière année

Plus de statistiques...

Timercan, Anatolie, Terriault, Patrick et Brailovski, Vladimir. 2023. « Axial tension/compression and torsional loading of diamond and gyroid lattice structures for biomedical implants: Simulation and experiment ». Materials and Design, vol. 225.
Compte des citations dans Scopus : 13.

[thumbnail of Terriault-P-2023-26150.pdf]
Prévisualisation
PDF
Terriault-P-2023-26150.pdf - Version publiée
Licence d'utilisation : Creative Commons CC BY.

Télécharger (9MB) | Prévisualisation

Résumé

Lattice structures are increasingly used in biomedical implants, notably intervertebral cages, requiring a better understanding of their behavior for the different types of loading they undergo during application. Strut-based diamond and sheet-based gyroid structures with porosity levels ranging from 50 to 80 % and an identical pore size of 750 μm were manufactured from Ti6Al4V alloy, tested experimentally and simulated numerically in axial tension/compression and in torsion to simulate flexion/extension, compression and rotation of the human spine. The manufactured structures were within 5 % of the targeted porosity. However, numerical simulations overestimated the experimental apparent (effective) stiffness and strength of the structures by an average of 25 %, likely due to the presence in them of manufacturing defects, especially in the higher porosity lattices. Experimental and numerical results showed that the structures have quasi-identical mechanical properties in compression and in tension. However, a comparison of the torsion and axial results indicated that conventional bulk material failure theories such as the von Mises limitation criterion do not apply to the apparent properties of lattice structures. Studied lattices exhibited adequate resistance for use in intervertebral cages, however their stiffness was greater than those of the vertebrae, while situated in the stiffness range of cortical bone.

Type de document: Article publié dans une revue, révisé par les pairs
Professeur:
Professeur
Terriault, Patrick
Brailovski, Vladimir
Affiliation: Génie mécanique, Génie mécanique
Date de dépôt: 26 janv. 2023 23:37
Dernière modification: 03 févr. 2023 16:51
URI: https://espace2.etsmtl.ca/id/eprint/26150

Actions (Authentification requise)

Dernière vérification avant le dépôt Dernière vérification avant le dépôt