Soyeux, Dan, Peixoto, Margaux, Tétreault, Patrice, Begon, Mickaël et Hagemeister, Nicola.
2025.
« Multiple acromion lengths and glenoid implant inclinations can result in the same critical shoulder angle with large differences in articular joint loading–a musculoskeletal study ».
JSES International.
(Sous presse)
Prévisualisation |
PDF
Hagemeister-N-2025-30686.pdf - Version acceptée Licence d'utilisation : Creative Commons CC BY-NC-ND. Télécharger (2MB) | Prévisualisation |
Résumé
Background: Glenoid implant loosening is the most common complication of anatomical total shoulder arthroplasty. It is caused by high glenohumeral shear forces and by an eccentric loading of the glenoid implant provoking its “rocking”. The critical shoulder angle (CSA) varies with the glenoid inclination and the acromion length. A higher CSA has been correlated with earlier radiological signs of glenoid loosening. However, the reliability of the CSA in predicting the risk factors of glenoid loosening has yet to be determined since the same CSA can result from multiple scapular anatomies. Methods: An inverse-dynamic musculoskeletal model in Anybody Modeling System of the shoulder with anatomical implants allowing glenohumeral translations was used. The acromion length and the glenoid implant inclination were varied to create multiple CSA configurations. Muscle forces, the force, and the moment applied to the glenoid implant were simulated during a shoulder abduction to compare the risks of glenoid loosening. Results: Increasing the CSA with an upward-tilted glenoid and a longer acromion led to more eccentric forces applied to the glenoid. The moment and shear applied to the glenoid implant increased with a higher CSA and were minimal for the smaller CSAs. Depending on the combination of inclination and acromion length, the shear and the moment were highly variable for the same CSA. Conclusion: Measuring the CSA as a global indicator may be insufficient to accurately predict the risk of glenoid loosening. It suggests that the acromion length could be considered during surgical planning to determine the adequate glenoid implant inclination.
Type de document: | Article publié dans une revue, révisé par les pairs |
---|---|
Professeur: | Professeur Hagemeister, Nicola |
Affiliation: | Génie des systèmes |
Date de dépôt: | 31 mars 2025 18:14 |
Dernière modification: | 10 avr. 2025 13:40 |
URI: | https://espace2.etsmtl.ca/id/eprint/30686 |
Actions (Authentification requise)
![]() |
Dernière vérification avant le dépôt |